3001 Levelogger 5 LTC, M100/C80
Levelogger Software includes a Calibration Wizard that guides the user through a 1 to 4-point calibration of the conductivity sensor, using standard solutions. The conductivity sensor autoranges from 0–100,000 µS/cm and calibrates from 50–80,000 µS/cm. Accuracy is ±1% from 5,000 µS/cm–80,000 µS/cm and the greater of ±2% or 15 µS/cm from 50 µS/cm–5,000 µS/cm.
The Levelogger 5 LTC can be set to record every 2 seconds to 99 hours, in linear, event-base, or scheduled sampling modes. It is compatible with all Levelogger communication and deployment accessories. A Barologger 5 can be used to compensate the water level readings.
Features:
Large memory: 100,000 sets of data
Stable communication: single-eye optical, easy to clean, more scratch resistant
Strong, robust design: double o-ring seals for two times over pressurization rating
Enhanced thermistor and conductivity sensitivity: accurate platinum RTD and conductivity sensor
Superior protection in harsh conditions: baked-on coating using polymerization—inside and out
Applications:
Salt water intrusion and soil salination monitoring
Plume remediation monitoring and studies
Leachate monitoring at landfills, mine tailings, waste disposal storage sites, and more
Agricultural and stormwater runoff monitoring
Create a historical database for potable water supply monitoring
Tracer tests
Specifications
Level Sensor: Ranges: Accuracy: Resolution: Units of Measure: Normalization: Temp. Comp. Range: |
Piezoresistive Silicon with Hastelloy Sensor 5, 10, 20, 30, 100, and 200 m ±0.05% FS 0.001% FS to 0.0006% FS cm, m, ft, psi, kPa, bar (ºC, ºF) Automatic Temperature Compensation 0º to 50ºC |
Temperature Sensor: Accuracy: Resolution: |
Platinum Resistance Temperature Detector (RTD) ±0.05ºC 0.003ºC |
Conductivity Sensor: Full Range: Calibrated Range: Accuracy: Resolution: Temp Comp. Range: Normalization: |
4-Electrode Platinum 0 – 100,000 µS/cm 50 – 80,000 µS/cm ±1%: 5,000 µS/cm – 80,000 µS/cm; greater of ±2% or 15 µS/cm: 50 µS/cm –5,000 µS/cm ±0.1 µS/cm 0ºC – 50ºC Specific Conductance @ 25˚C |
Battery Life: Clock Accuracy (typical): Operating Temperature: |
8 Years (1 reading every 5 minutes) ±1 minute/year (-20ºC to 80ºC) -20ºC to 80ºC |
Maximum Readings: Memory: Communication: |
100,000 sets of readings Slate or Continuous Optical high-speed: 57,600 bps with USB |
Size: Weight: Corrosion Resistance Wetted Materials: |
22 mm x 208 mm (7/8" x 8.2") 200 grams (7.06 oz) Baked-on coating using polymerization Platinum, Delrin®, Viton®, 316L Stainless Steel, Hastelloy, Regulator approved PFAS-free PTFE (inside and out) |
Sampling Mode: Measurement Rates: Barometric Compensation: |
Linear, Event & User-Selectable with Repeat Mode, Future Start, Future Stop, Real-Time View 2 seconds to 99 hours Software Wizard and Barologger 5 |